| Instrument |
Measuring Range |
| EFL |
Dioptric Power |
BFL and Radius |
MTF |
OptiSpheric IOL
OS IOL Tray TT
OS IOL Tray SA
|
+3 .....+500 mm
-3 ..... -500 mm |
+5 .....+300 D
-5 ..... -300 D |
±3 .....±400mm |
absolute MTF-Test: 5...75 mm EFL
relative MTF-Test : 5-500 mm EFL |
OptiSpheric IOL LR
OS IOL LR Tray TT
OS IOL LR Tray SA
|
+3 .....+1000 mm
-3 ..... -1000 mm |
+1 .....+300 D
-1 ..... -300 D |
±5 .....±500mm |
absolute MTF-Test: 5...150 mm EFL
relative MTF-Test : 5-1000 mm |
| System Performance |
EFL |
| Resolution: |
0.03...0.2% |
| Measurement Accuracy |
5...25 mm: . . . . . . . . . . . . . . . . . . . . . . . .0.1%...0.3%
25...500 mm: . . . . . . . . . . . . . . . . . . . . . .0.03%...0.1%
500...1000 mm: . . . . . . . . . . . . . . . . . . .a 0.05%...0.3% |
| Measurement Time EFL |
Autofocus: First measurement:5-6 sec
Next measurements: 3-5 sec |
| |
Power |
| Repeatability: |
0.03...0.2% |
| Measurement Accuracy |
5...40 Diopter: . . . . . . . . . . . . . . . . . . . . . .+/-0.1%
40...200 Diopter: . . . . . . . . . . . . . . . . . . . .+/-0.12%
>200 Diopter: . . . . . . . . . . . . . . . . . . . . . .+/-0.2% |
| Measurement Time Dioptric Power |
Autofocus: First measurement:5-6 sec
Next measurements: 3-5 sec |
Measurement Time for complete IOL
characterization EFL, Power, MTF |
Monofocal: < 15 sec. |
Multifocal: < 45 sec. |
| |
MTF |
| Wavelength range |
400-1000nm |
| Spatial frequency: |
0-500 lp/mm-higher frequencies available on request |
| |
MTF-Measurement / Slit evaluation |
| Repeatability: |
±0.01 MTF |
| Accuracy |
±0.02 MTF |
| |
BFL, FFL and Radius of curvature |
| Repeatability: |
0.02 to 0.2% |
| Accuracy: |
0.03 to 0.3% |