Description
OptiSpheric® IOL set up follows the requirements of the ISO
11979 standard for the different parameters to be measured.
An illuminated target is at the focus of a collimator, so that
parallel light is incident upon the IOL.
All targets needed for full characterisation of the IOL are
mounted on a fully automated reticle changer.
Double slits are used for power determination, single slit, cross
or pinhole are used for MTF measurement.
The collimator is an achromat that is virtually free of
aberrations for the wavelength band transmitted by the filter
(narrow band green filter centered at 546 nm).
A microscope with a high numerical aperture allows for imaging on a high
resolution CCD-Camera connected to a frame grabber mounted in a PC.
The complete measurement head, including the microscope and the
sensor, is mounted on a motorized stage featuring micro step
operation for fast and
accurate autofocus. Low noise cameras for demanding IOLs can also
be implemented in the instrument.
|